HeartSine Samaritan® PAD - Public Access Defibrillator |
*Self Compensating Output Pulse Envelope technology automatically compensates energy, slope and pulse envelope for the patient. **Efficacy of Distinct Energy Delivery Protocols Comparing Two Biphasic Defibrillators for Cardiac Arrest, Walsh, McClelland, Owens, Anderson, Turner, Adgey; The American Journal of Cardiology, Vol. 94, Aug. 1, 2004 |
CAUTION: U.S. Federal law restricts this device to sale by or on the order of a licensed practitioner. |
![]() |
|
Technical Overview |
|
Physical |
With Pad-Pak™ Battery Inserted |
Size: |
8.0 in x 7.25 in x 1.9 in (20cm x 18.4cm x 4.8cm) |
Weight: |
2.4lbs (1.1 kg) including Pad-Pak™ Battery |
Defibrillator |
|
Waveform: |
(Self-Compensating Output Pulse Envelope) Biphasic waveform. Optimized biphasic escalating waveform compensates energy, slope and envelope for patient impedance |
Energy Selection |
Pre-configured factory settings for escalating energy are AHA/ERC 2005 |
Adult |
|
Pediatric |
|
Charging Time |
|
New Battery: |
Typically 150J in <8 sec., 200J in < 12 sec. |
After 6 discharges: |
Typically 150J in <8 sec., 200J in < 12 sec. |
Patient Analysis System |
|
Method: |
Evaluates patient's ECG, signal quality, electrode contact integrity and patient impedance to determine if defibrillation is required |
Sensitivity/Specificity: |
Meets ISO 60601-2-4 and AAMI DF80:2003. |
Environmental |
|
Operating/Standby |
0°C to 50°C (+32°F to + 122°F) |
Temporary Transportation |
-10°C to 50°C (14°F to 122°F) for up to two days. Unit must be returned to standby/operating temperature for 24 hours before use. |
Relative Humidity: |
5% to 95% (non-condensing) |
Water Resistance: |
IEC 60529/EN 60529 IP56 |
Altitude: |
0 to 15,000 feet (0 - 4,575 meters) |
Shock: |
MIL STD 810F Method 516.5, Procedure I (40G's) |
Vibration: |
MIL STD 810F Method 514.5+ |
Category 4 Truck Transportation - US Highways |
|
Category 7 Aircraft - Jet 737 & General Aviation (Exposure) |
|
EMC: |
EN 60601-1-2, Second Edition: 2002 |
Radiated Emissions: |
ENSS-11 :1999+A2:2001 |
Electrostatic Discharge |
EN61000-4-3:2001 80MHZ-2.5GHZ (10V/m) |
Magnetic Field Immunity: |
EN61000-4-8:2001 (3 A/m) |
Aircraft: |
RTCA / DO -1600: 1997, Section 21 (Category M) TSO-C142/RTCA 00-227 |
Falling height: |
1 meter |
Event Documentation |
|
Type: |
Internal memory |
Memory Capacity: |
45 minutes of ECG (full disclosure) and event! incident recording |
Playback Capabilities: |
Custom USB cable directly connected to PC and Saver™ EVO Windows-based data review software |
Materials Used |
|
PAD SAM300P: |
ABS, Santoprene. Printed circuit board with electronic components. |
PAD Cartridge: |
Battery: Lithium Manganese Dioxide |
Housing: |
ABS - Electrodes: Hydrogel, Silver, Aluminium and Polyester |
Pad-Pak™ - Electrode and Battery Cartridge |
|
Adult Pad-Pak (Pad-Pak-01) and Pediatric Pad-Pak (Pad-Pak-02) |
|
Shelf Life: |
Typically 3.5 years from manufacture date |
Weight: |
0.44 lbs (0.2kg) |
Size: |
3.93 in x 5.24 in x .94 in (10cm x 13.3cm x 2.4cm) |
Battery Type: |
Lithium Manganese Dioxide (LiMnO2) 18V, 0.8 Amp Hrs |
Capacity: |
>30 shocks at 200J or 6 hours of continuous monitoring |
Electrodes: |
Samaritan® disposable defibrillation pads are - supplied as standard with each device |
Placement: |
Anterior-lateral (Adult); |
Anterior-posterior (Pediatric) |
|
Active Gel Area: |
100 cm2 |
Cable Length: |
3.5 ft (1m) |
WSP home page | Products page | Suppliers page |